FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY PLUS

MDR report key: 2812308 · Received October 31, 2012

Report

Report Number
2649622-2012-16670
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 1, 2012
Report Date
October 1, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006/S002
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THERE WAS BLOOD (NOT OBSTRUCTED) ON THE DISTAL AND PROXIMAL CONDUCTORS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT WHILE MANIPULATING THE LEAD AND ANGIOPLASTY WIRE, IT WAS NOTED THE WIRE NO LONGER PROTRUDED THROUGH THE TIP SEAL OF THE LEAD. THE WIRE WAS BOUND AT THE TIP OF THE LEAD AND WOULD NOT EXTEND. THE LEAD WAS NOT USED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY PLUS DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MEDTRONIC PUERTO RICO, INC. 4296

Patients

Seq Age Sex Outcome Treatment
1 Other