FDA Adverse Event Injury Summary report: N

PULSE-GENERATOR, PACEMAKER, IMPLANTABLE

MDR report key: 2812234 · Received October 31, 2012

Report

Report Number
2182208-2012-03743
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
DTB
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. NO ANOMALIES WERE FOUND FROM THE ANALYSIS OF THE DEVICE MEMORY.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. WITHOUT A LOT NUMBER OR DEVICE SERIAL NUMBER, THE MANUFACTURING DATE CANNOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS OVERSENSING MYOPOTENTIALS AND INHIBITING PACING. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DTB MEDTRONIC, INC. LEAD/MEDT

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R T70A2 IMPLANTABLE PULSE GENERATOR