FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR

MDR report key: 2812221 · Received October 31, 2012

Report

Report Number
6000144-2012-06346
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE DEVICE MET 83% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE BATTERY DID NOT LAST AS EXPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. C154DWK

Patients

Seq Age Sex Outcome Treatment
1 67 YR Hospitalization| R 5568 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD