FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2812049 · Received October 31, 2012

Report

Report Number
2649622-2012-16650
Event Type
Injury
Date Received
October 31, 2012
Date of Event
September 25, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THE OUTER INSULATION HAD A COSMETIC DEPRESSION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SCHEDULED DEVICE CHANGE-OUT, THE RIGHT VENTRICULAR [RV] WAS INADVERTENTLY CUT DURING POCKET EVACUATION. THE RV LEAD WAS PARTIALLY EXPLANTED AND REPLACED; THE PORTION OF THE LEAD REMAINING IN THE PATIENT WAS INACTIVATED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 60 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB