FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 2812016
·
Received October 31, 2012
Report
- Report Number
- 2649622-2012-16156
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- September 19, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR: (B)(4) - THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT THE PHYSICIAN FELT THAT THE HELIX WOULD NOT FULLY RETRACT. THE HELIX WAS EXTENDED INTO THE TISSUE; THE PHYSICIAN WAS TRYING TO GET A GOOD POSITION. WHEN THE PHYSICIAN ATTEMPTED TO RETRACT THE HELIX, THE LEAD DISLODGED. A DIFFERENT LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Other |