FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2812007 · Received October 31, 2012

Report

Report Number
2649622-2012-17402
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4): THE FULL LEAD WAS RETURNED TO THE MANUFACTURER AND ANALYZED AND NO ANOMALIES WERE FOUND. THE PROXIMAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED) AND WAS DISTORTED. THE OUTER INSULATION HAD COSMETIC CUT, WAS BREACHED CUT AND HAD COSMETIC DEPRESSION. THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. THE LEAD HAD APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

THE RIGHT VENTRICULAR (RV) LEAD WAS RETURNED TO THE MANUFACTURER DUE TO SYSTEM UPGRADE, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI

Patients

Seq Age Sex Outcome Treatment
1 84 YR Hospitalization| R (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR