FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 2812002 · Received October 31, 2012

Report

Report Number
6000144-2012-05979
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Removal / Correction Number
Z-0111-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE DEVICE WAS RETURNED, ANALYZED, AND ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION, MEETING 80% OF THE EXPECTED LONGEVITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) REACHED EARLY END OF SERVICE (EOS). THE ICD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSULTA CRT-D IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. D224TRK

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization| R 6944 IMPLANTABLE TACHY LEAD| 4195 IMPLANTABLE PACING LEAD| 4076 IMPLANTABLE PACING LEAD