FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2811996 · Received October 31, 2012

Report

Report Number
2649622-2012-16411
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SCHEDULED DEVICE CHANGE-OUT PROCEDURE, A PRE-EXISTING "NICK" WAS SEEN IN THE HIGH VOLTAGE PORTION OF THE RIGHT VENTRICULAR [RV] LEAD. THE LEAD WAS ASSESSED AS FUNCTIONAL AS ALL MEASUREMENTS WERE NORMAL AND THE NICK WAS REPAIRED. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization| R 7230CX IMPLANTABLE PACEMAKER/CARDIO/DEFIB