- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY- THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES FOUND. THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THE OVERLAY TUBING HAD COSMETIC ENVIRONMENTAL STRESS CRACKING (ESC).
Description of Event or Problem · 1
IT WAS REPORTED THAT THERE WAS AN APPARENT INSULATION BREACH. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT FIDELIS | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6949 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4194 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD |