FDA Adverse Event Injury Summary report: N

SPRINT FIDELIS

MDR report key: 2811900 · Received October 31, 2012

Report

Report Number
2649622-2012-16885
Event Type
Injury
Date Received
October 31, 2012
Report Date
January 2, 2013
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015/S029
Removal / Correction Number
Z-0070-2008
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY- THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES FOUND. THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THE OVERLAY TUBING HAD COSMETIC ENVIRONMENTAL STRESS CRACKING (ESC).

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS AN APPARENT INSULATION BREACH. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT FIDELIS DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6949

Patients

Seq Age Sex Outcome Treatment
1 83 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4194 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD