FDA Adverse Event Malfunction Summary report: N

INSYNC III MARQUIS

MDR report key: 2811853 · Received October 31, 2012

Report

Report Number
6000094-2012-02473
Event Type
Malfunction
Date Received
October 31, 2012
Manufacturer
MEDTRONIC S.A.
Product Code
NIK
PMA / PMN Number
P010031/S018
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 84% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. THE BATTERY VOLTAGE IS PRE/APPROACHING ELECTIVE REPLACEMENT INDICATOR (ERI). WEEKLY BATTERY VOLTAGE TREND DATA IN SAVE TO DISK FILE (B)(4) SHOWS MIN B(V) =2.92 TO 2.64 VOLTS BETWEEN (B)(4) 2012 WHICH IS BEFORE DEVICE ERI <= 2.62 VOLT.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE SHOWED A SUDDEN DROP IN BATTERY VOLTAGE IN A SHORT AMOUNT OF TIME. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE DEVICE WAS AT ELECTIVE REPLACEMENT INDICATOR (ERI) AND WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSYNC III MARQUIS DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC S.A. 7279

Patients

Seq Age Sex Outcome Treatment
1 56 YR Hospitalization| O| R 4076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| 6935 IMPLANTABLE TACHY LEAD