FDA Adverse Event Injury Summary report: N

CAPSURE Z NOVUS

MDR report key: 2811833 · Received October 31, 2012

Report

Report Number
2649622-2012-16881
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P850089/S38
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD HIGH IMPEDANCE AND HIGH THRESHOLD. THE LEAD WAS CAPPED AND WAS REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS WERE NOTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE Z NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5054

Patients

Seq Age Sex Outcome Treatment
1 79 YR Hospitalization| R (B)(4) IMPLANTABLE PULSE GENERATOR