FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2811718 · Received October 31, 2012

Report

Report Number
2649622-2012-15884
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
September 7, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVY
PMA / PMN Number
P920015/S055
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED. THE DISTAL CONDUCTOR WAS DISTORTED AND THE DEFIBRILLATION CONDUCTOR WAS CUT. THERE WAS BLOOD/BODY FLUID (NOT OBSTRUCTED) ON ALL CONDUCTORS. THE OUTER INSULATION WAS BREACHED CUT. THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. VISUAL ANALYSIS REVEALED THE LEAD WAS DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE LEAD REQUIRED SEVERAL TURNS FOR EXTENSION AND RETRACTION OF THE HELIX. IT WAS ALSO NOTED THE HELIX WOULD "POP" OUT RATHER THAN GRADUALLY EXTEND. SEVERAL IMPLANT ATTEMPTS WERE MADE WITH THE LEAD WITHOUT SUCCESS. THE LEAD WAS NOT USED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD NVY MEDTRONIC PUERTO RICO, INC. 6947M

Patients

Seq Age Sex Outcome Treatment
1 Other