ATTAIN ABILITY PLUS

MDR report key: 2811658 · Received October 31, 2012

Report

Report Number
2649622-2012-16868
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 1, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006/S002
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED) AND THE LEAD APPEARED DAMAGED AT IMPLANT. VISUAL ANALYSIS NOTED THAT THE LEAD WAS FLUSHED CONTRARY TO RECOMMENDED PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, WHEN THE LEFT VENTRICULAR LEAD WAS FLUSHED, THE RINSE WATER CAME OUT THROUGH THE LEAD INSULATION. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY PLUS IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4296

Patients

Seq Age Sex Outcome Treatment
1 Other