FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 2811627
·
Received October 31, 2012
Report
- Report Number
- 2649622-2012-16865
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- October 5, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE DEVICE WAS RETURNED AND ANALYSIS FOUND NO ANOMALIES. HOWEVER, A SET SCREW WAS LOOSE/DETACHED.
Description of Event or Problem · 1
IT WAS REPORTED THAT AT A ROUTINE DEVICE CHANGE OUT, IT WAS DISCOVERED THAT THE ATRIAL LEAD WAS DISLODGED. THE LEAD WAS CAPPED AND REPLACED. WHEN THE PHYSICIAN CONNECTED THE NEW ATRIAL LEAD TO THE NEW DEVICE, THE PHYSICIAN WAS UNABLE TO ENGAGE THE SETSCREW OF THE NEW DEVICE. THE DEVICE WAS NOT IMPLANTED AND ANOTHER DEVICE WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Hospitalization| R | (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD |