CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS
Report
- Report Number
- 2015691-2012-18583
- Event Type
- Injury
- Date Received
- October 31, 2012
- Date of Event
- September 18, 2012
- Report Date
- October 5, 2012
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- DYE
- PMA / PMN Number
- P860057/S001
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- OTHER
Narratives
DEVICE NOT RETURNED. DESPITE ATTEMPTS TO RECEIVE FURTHER INFORMATION REGARDING THE DEVICE AND EVENT, NO ADDITIONAL INFORMATION OR SAMPLE FOR EVALUATION WAS RECEIVED. THEREFORE, THE ROOT CAUSE OF THIS EVENT COULD NOT BE CONFIRMED. THERE HAVE NOT BEEN ANY ALLEGATIONS OF A MALFUNCTION OR DEFICIENCY RELATED TO THE EXPLANTED VALVE. CONTINUED ATTEMPTS ARE BEING MADE TO OBTAIN ADDITIONAL DETAILS REGARDING THIS CASE. A SUPPLEMENTAL MDR WILL BE SUBMITTED IN THE EVENT ANY NEW INFORMATION IS RECEIVED.
ADDITIONAL MANUFACTURER NARRATIVE: BASED THE OPERATIVE REPORT PROVIDED, THE EXPLANT WAS DUE TO AORTIC VALVE STENOSIS. THE PROSTHETIC VALVE LEAFLETS WERE COMPLETELY CALCIFIED. THE ASCENDING AORTA MEASURED 4.8-5CM BY TEE. THE EF HAD IMPROVED FROM 30% IMMEDIATELY PERIOPERATIVELY TO 40-45% POST-IMPLANT. THERE ARE SEVERAL POTENTIAL CAUSES FOR PROSTHETIC VALVE STENOSIS (E.G. CALCIFICATION, PANNUS GROWTH). UNFORTUNATELY, THE VALVE WAS NOT RETURNED TO EDWARDS FOR ANALYSIS, AND THEREFORE, THE ROOT CAUSE OF THIS EVENT COULD NOT BE INVESTIGATED. ALTHOUGH THE ROOT CAUSE CANNOT BE INVESTIGATED, THE REPORTED STENOSIS WAS LIKELY DUE TO THE "PROSTHETIC VALVE LEAFLETS WERE COMPLETELY CALCIFIED" NOTED IN THE OPERATIVE REPORT. CALCIFICATION IS A WELL RECOGNIZED FAILURE MODE OF BIOPROSTHETIC VALVES. THE MECHANISMS FOR BIOPROSTHETIC HEART VALVE TISSUE CALCIFICATION ARE NOT FULLY UNDERSTOOD. MANY FACTORS CAN CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION INCLUDING PATIENT RELATED (E.G. PATIENT AGE, DISEASE STATE, IMMUNE STATUS, AND OTHER CO-MORBIDITIES), PHARMACOLOGICAL, AND INTRINSIC PROPERTIES OF THE VALVE ITSELF. IT IS WIDELY UNDERSTOOD THAT PATIENTS WITH CHRONIC RENAL DISEASE AND PRIOR HISTORY OF CALCIFIC STENOSIS OF THE NATIVE VALVE MAY BE PREDISPOSED TO BIOPROSTHETIC CALCIFICATION. NO FURTHER ACTIONS ARE POSSIBLE WITH THE AVAILABLE INFORMATION.
IN THIS CASE, IT WAS REPORTED THAT THE AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF APPROXIMATELY 9 YEARS. THE REASON FOR EXPLANT IS UNKNOWN. DESPITE ATTEMPTS TO RECEIVE FURTHER INFORMATION REGARDING THE DEVICE AND EVENT, NO ADDITIONAL INFORMATION OR SAMPLE FOR EVALUATION WAS RECEIVED. THERE HAVE NOT BEEN ANY ALLEGATIONS OF A MALFUNCTION OR DEFICIENCY RELATED TO THE EXPLANTED VALVE. THE DEVICE WAS REPLACED WITH ANOTHER EDWARDS BIOPROSHETIC VALVE. NO OTHER DETAILS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS | REPLACEMENT HEART VALVE | DYE | EDWARDS LIFESCIENCES | 2800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Hospitalization| R |