FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2811554 · Received October 31, 2012

Report

Report Number
2649622-2012-16384
Event Type
Injury
Date Received
October 31, 2012
Report Date
November 13, 2012
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015/S24
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION, HOWEVER PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND ANALYZED. ANALYSIS REVEALED NO ANOMALIES.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCKS DUE TO T-WAVE OVERSENSING (TWOS) OF THE VENTRICULAR LEAD. THE SENSITIVITY ON THE DEVICE WAS REPROGRAMMED AND THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCKS DUE TO T-WAVE OVERSENSING (TWOS) OF THE VENTRICULAR LEAD. THE SENSITIVITY ON THE DEVICE WAS REPROGRAMMED AND THE LEAD REMAINS IN U SE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6947

Patients

Seq Age Sex Outcome Treatment
1 48 YR Life Threatening| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD