FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 2811490
·
Received October 31, 2012
Report
- Report Number
- 2649622-2012-15865
- Event Type
- Injury
- Date Received
- October 31, 2012
- Date of Event
- September 1, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE RIGHT VENTRICULAR LEAD IMPEDANCE WAS HIGH AND INCREASING, HOWEVER BECAUSE OF THE GOOD THRESHOLD AND SENSING, THE LEAD WAS IMPLANTED. THE FOLLOWING DAY, HIGH IMPEDANCE WAS NOTED AGAIN AND ADDITIONALLY, THE THRESHOLD WAS VARIABLE AND HIGH. THE LEAD WAS REPOSITIONED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| R | (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD |