FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 2811486 · Received October 31, 2012

Report

Report Number
6000144-2012-05961
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Removal / Correction Number
Z-0111-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4): THE DEVICE WAS RETURNED TO THE MANUFACTURER ANA ANALYZED. THE DEVICE MET 97% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE REACHED EARLY BATTERY DEPLETION. IT WAS ALSO REPORTED THAT THE DEVICE REACHED RECOMMENDED REPLACEMENT TIME (RRT) BEFORE THREE YEARS. THEREFORE THE DEVICE WAS REMOVED AND REPLACED WITH A NEW DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSULTA CRT-D IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. D224TRK

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R 4196 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD