FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 2811482 · Received October 31, 2012

Report

Report Number
6000144-2012-05823
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S062
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. FOLLOW-UP INFORMATION RECEIVED FROM THE CLINIC DISCLOSED THAT THE DEVICE WAS EXPLANTED DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154AWG

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R 5072 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD