FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2811393 · Received October 31, 2012

Report

Report Number
2649622-2012-15858
Event Type
Injury
Date Received
October 31, 2012
Date of Event
May 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND NO ANOMALIES WERE FOUND. HIGH THRESHOLDS WERE NOTED IN THE SAVE TO DISK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED DIAPHRAGMATIC STIMULATION AFTER THE ATRIAL LEAD DISLODGED FROM THE PREVIOUS POSITION AND PULLED BACK UP INTO THE SUPERIOR VENA CAVA AND RIGHT ATRIAL JUNCTION. THERE WERE ALSO HIGH/UNSTABLE THRESHOLDS. THE LEAD WAS REPOSITIONED AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC PUERTO RICO, INC. 5076

Patients

Seq Age Sex Outcome Treatment
1 56 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR