FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 2811346 · Received October 31, 2012

Report

Report Number
2649622-2012-16109
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
September 6, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4). THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THE HELIX/LOBE WAS DISTORTED/BENT, BLOOD WAS FOUND IN/ON THE HELIX/LOBE MECHANISM AND THE HELIX/LOBE MECHANISM (SLEEVE HEAD), AND THE LEAD APPEARED TO HAVE BEEN DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE LEAD EXHIBITED HIGH IMPEDANCES AND HIGH THRESHOLDS. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE S IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6935

Patients

Seq Age Sex Outcome Treatment
1 Other