SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2012-16109
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- September 6, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4). THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THE HELIX/LOBE WAS DISTORTED/BENT, BLOOD WAS FOUND IN/ON THE HELIX/LOBE MECHANISM AND THE HELIX/LOBE MECHANISM (SLEEVE HEAD), AND THE LEAD APPEARED TO HAVE BEEN DAMAGED AT IMPLANT.
IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE LEAD EXHIBITED HIGH IMPEDANCES AND HIGH THRESHOLDS. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |