FDA Adverse Event
Injury
Summary report: N
CAPSURE Z NOVUS
MDR report key: 2811343
·
Received October 31, 2012
Report
- Report Number
- 2649622-2012-16361
- Event Type
- Injury
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P850089/S38
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD OVERSENSING AND THE IMPEDANCE WAS VARYING. THE DEVICE WAS REPROGRAMMED AND THE LEAD REMAINS IMPLANTED BUT ELECTRICALLY ABANDONED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE Z NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5054 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Required Intervention | P1501DR IMPLANTABLE PULSE GENERATOR| 5554 IMPLANTABLE PACING LEAD |