FDA Adverse Event Malfunction Summary report: N

CARELINK PROGRAM W/TELEMETRY C

MDR report key: 2811245 · Received October 31, 2012

Report

Report Number
2182208-2012-03756
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
September 28, 2012
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003/S080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS FOUND THAT THE SCREEN HAD COLORED LINES AND REMAINED BLANK. THE FLEX TAPE WAS MISALIGNED TO THE LOW VOLTAGE DIFFERENTIAL SIGNALING (LVDS).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER WAS UNABLE TO BOOT UP TO THE MAIN SCREEN, THAT THE SCREEN REMAINED BLACK. THE PROGRAMMER WAS RETURNED FOR SERVICE. IT WAS INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAM W/TELEMETRY C PROGRAMMER DXY MEDTRONIC, INC. 2090W

Patients

Seq Age Sex Outcome Treatment
1 Other 2067 PROGRAMMER RF HEAD| 2290 PACING SYSTEM ANALYZER