FDA Adverse Event
Malfunction
Summary report: N
CARELINK PROGRAM W/TELEMETRY C
MDR report key: 2811245
·
Received October 31, 2012
Report
- Report Number
- 2182208-2012-03756
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- September 28, 2012
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DXY
- PMA / PMN Number
- P890003/S080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS FOUND THAT THE SCREEN HAD COLORED LINES AND REMAINED BLANK. THE FLEX TAPE WAS MISALIGNED TO THE LOW VOLTAGE DIFFERENTIAL SIGNALING (LVDS).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PROGRAMMER WAS UNABLE TO BOOT UP TO THE MAIN SCREEN, THAT THE SCREEN REMAINED BLACK. THE PROGRAMMER WAS RETURNED FOR SERVICE. IT WAS INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK PROGRAM W/TELEMETRY C | PROGRAMMER | DXY | MEDTRONIC, INC. | 2090W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | 2067 PROGRAMMER RF HEAD| 2290 PACING SYSTEM ANALYZER |