FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2811191 · Received October 31, 2012

Report

Report Number
2649622-2012-16818
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
September 27, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE INNER INSULATION WAS KINKED/BUCKLED, THE INNER TUBING WAS KINKED/BUCKLED, THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM, THE LEAD WAS STRETCHED, AND VISUAL ANALYSIS REVEALED THE LEAD WAS DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT ATTEMPT THE RIGHT ATRIAL (RA) LEAD HAD HIGH IMPEDANCE MEASUREMENTS IN MULTIPLE LOCATIONS WITHIN THE RIGHT ATRIUM. THE RA LEAD WAS NOT USED AND A NEW LEAD WAS IMPLANTED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI

Patients

Seq Age Sex Outcome Treatment
1 62 YR Other