FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS MRI
MDR report key: 2811191
·
Received October 31, 2012
Report
- Report Number
- 2649622-2012-16818
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- September 27, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE INNER INSULATION WAS KINKED/BUCKLED, THE INNER TUBING WAS KINKED/BUCKLED, THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM, THE LEAD WAS STRETCHED, AND VISUAL ANALYSIS REVEALED THE LEAD WAS DAMAGED AT IMPLANT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT ATTEMPT THE RIGHT ATRIAL (RA) LEAD HAD HIGH IMPEDANCE MEASUREMENTS IN MULTIPLE LOCATIONS WITHIN THE RIGHT ATRIUM. THE RA LEAD WAS NOT USED AND A NEW LEAD WAS IMPLANTED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 5086MRI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Other |