FDA Adverse Event
Injury
Summary report: N
ENRHYTHM DR
MDR report key: 2811169
·
Received October 31, 2012
Report
- Report Number
- 6000144-2012-06070
- Event Type
- Injury
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S38
- Removal / Correction Number
- Z-1440-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. POWER ON RESET - POWER ON RESET PARAMETERS. DEVICE EXPERIENCED A CRITICAL RAM PARITY ERROR POR ON (B)(6) 2012 IN LOCATION "18 A2." DEVICE SHOULD BE ABLE TO RECOVER FROM THE EVENT AND CONTINUE NORMAL FUNCTION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE EXPERIENCED AN ELECTRICAL RESET. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENRHYTHM DR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC MED REL, INC. | P1501DR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 90 YR | Required Intervention | 5076 X2 IMPLANTABLE PACING LEAD |