FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2811081 · Received October 31, 2012

Report

Report Number
2649622-2012-15828
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THERE WERE NO ANOMALIES FOUND.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE LEAD WAS EXPLANTED RATHER THAN CAPPED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE ALERTED FOR HIGH RIGHT VENTRICULAR IMPEDANCE. A LEAD INTEGRITY ALERT ALSO TRIGGERED. THE LEAD HAD BEEN DESTROYED DUE TO TWIDDLER'S SYNDROME. THE LEAD WAS CAPPED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 85 YR Hospitalization| R D224VRC IMPLANTABLE PACEMAKER/CARDIO/DEFIB