FDA Adverse Event Injury Summary report: N

VITATRON T60 DR

MDR report key: 2810919 · Received October 31, 2012

Report

Report Number
6000094-2012-02459
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P990001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ES
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DEVICE INTERROGATION A WARNING APPEARED STATING "TO DO A SAVE TO DISK (STD) AND CONTACT THE REPRESENTATIVE". ADDITIONAL INFORMATION NOTED THAT THE DEVICE HAD A BIT FLIP AND A POWER ON RESET (POR). A MANUAL GUIDED RESET (MGR) WILL BE PERFORMED TO REPROGRAM THE DEVICE, HOWEVER, THIS HAS NOT YET OCCURRED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT A MANUAL GUIDED RESET (MGR) HAS TAKEN PLACE AND THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITATRON T60 DR PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY MEDTRONIC S.A. T60A2

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention