FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 2810890
·
Received October 31, 2012
Report
- Report Number
- 2649622-2012-15813
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- September 3, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVY
- PMA / PMN Number
- P920015/S055
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, R-WAVES THROUGH THE ANALYZER WERE AT 8.0 MV, BUT THROUGH THE DEVICE THEY WERE LESS THAN 3.0 MV. THE LEAD WAS NOT USED AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | NVY | MEDTRONIC PUERTO RICO, INC. | 6947M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Other | D314DRM IMPLANTABLE PACEMAKER/CARDIO/DEFIB |