FDA Adverse Event Injury Summary report: N

INTRINSIC

MDR report key: 2810838 · Received October 31, 2012

Report

Report Number
6000094-2012-02371
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC S.A.
Product Code
LWS
PMA / PMN Number
P980016/S041
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. POWER ON RESET PARAMETERS (POR) WERE NOTED AS ONE POR FOR WRITE TO LOCKED RAM, ADDR=0DC4, DATA=1 OCCURRED ON (B)(6) 2012. ONE PATIENT ALERT FOR POR OCCURRED ON (B)(6) 2012.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD AN ELECTRICAL RESET. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTRINSIC IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC S.A. 7288

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention 6940 IMPLANTABLE PACING LEAD| 6943 IMPLANTABLE TACHY LEAD