FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 2810758 · Received October 31, 2012

Report

Report Number
2649622-2012-16043
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THE PROXIMAL CONDUCTOR WAS DISTORTED, AND THE OUTER INSULATION WAS MELTED AND BREACHED CUT, AND EXHIBITED COSMETIC ENVIRONMENTAL STRESS CRACKING.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD WAS CAUSING STIMULATION OF THE DIAPHRAGM AND HAD HIGH THRESHOLDS. THE LV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4194

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization| R 8042 IMPLANTABLE PULSE GENERATOR| 5076 X2 IMPLANTABLE PACING LEADS