FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2810743 · Received October 31, 2012

Report

Report Number
2649622-2012-16300
Event Type
Injury
Date Received
October 31, 2012
Date of Event
February 10, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED AN INAPPROPRIATE DUE TO TWAVE OVERSENSING (TWOS). ADJUSTING THE RIGHT VENTRICULAR LEAD SENSITIVITY GOT RID OF THE TWOS. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 46 YR Life Threatening| R D274DRG IMPLANTABLE PACEMAKER/CARDIO/DEFIB