FDA Adverse Event
Injury
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 2810743
·
Received October 31, 2012
Report
- Report Number
- 2649622-2012-16300
- Event Type
- Injury
- Date Received
- October 31, 2012
- Date of Event
- February 10, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT RECEIVED AN INAPPROPRIATE DUE TO TWAVE OVERSENSING (TWOS). ADJUSTING THE RIGHT VENTRICULAR LEAD SENSITIVITY GOT RID OF THE TWOS. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Life Threatening| R | D274DRG IMPLANTABLE PACEMAKER/CARDIO/DEFIB |