FDA Adverse Event Injury Summary report: N

ENRHYTHM MRI

MDR report key: 2810712 · Received October 31, 2012

Report

Report Number
6000094-2012-02370
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC S.A.
Product Code
LWP
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE EXPERIENCED POSSIBLE PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. THE PATIENT IS ENROLLED IN THE (B)(4) CLINICAL STUDY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM MRI IMPLANTABLE PULSE GENERATOR LWP MEDTRONIC S.A. EMDR01

Patients

Seq Age Sex Outcome Treatment
1 83 YR Hospitalization| R (B)(4) X2 IMPLANTABLE PACING LEADS