FDA Adverse Event
Injury
Summary report: N
ENRHYTHM MRI
MDR report key: 2810712
·
Received October 31, 2012
Report
- Report Number
- 6000094-2012-02370
- Event Type
- Injury
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- LWP
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE EXPERIENCED POSSIBLE PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. THE PATIENT IS ENROLLED IN THE (B)(4) CLINICAL STUDY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENRHYTHM MRI | IMPLANTABLE PULSE GENERATOR | LWP | MEDTRONIC S.A. | EMDR01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Hospitalization| R | (B)(4) X2 IMPLANTABLE PACING LEADS |