FDA Adverse Event Injury Summary report: N

CAPSURE SENSE

MDR report key: 2810613 · Received October 31, 2012

Report

Report Number
2649622-2012-15772
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P830061/S034
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE ACTUAL LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. IMPEDANCE/HIGH IMPEDANCE: V-LEAD IMPEDANCE TREND BEGINS A LINEAR RISE OVER 17 WEEKS BEFORE LEAD WAS REVISED. THRESHOLDS: THRESHOLDS INCREASE OVER 16 WEEKS, STARTING 27-MAY, TOPPING OUT AT 1V. WITHIN NORMAL RANGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD EXHIBITED SLOWLY INCREASING THRESHOLDS AND IMPEDANCES, REACHING HIGH LEVELS OF IMPEDANCE. IT WAS FURTHER REPORTED THAT THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD EXHIBITED SLOWLY INCREASING THRESHOLDS AND IMPEDANCES, REACHING HIGH LEVELS OF IMPEDANCE. IT WAS FURTHER REPORTED THAT THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SENSE ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC PUERTO RICO, INC. 4074

Patients

Seq Age Sex Outcome Treatment
1 59 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB