CAPSURE SENSE
Report
- Report Number
- 2649622-2012-15772
- Event Type
- Injury
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P830061/S034
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
PRODUCT EVENT SUMMARY: THE ACTUAL LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. IMPEDANCE/HIGH IMPEDANCE: V-LEAD IMPEDANCE TREND BEGINS A LINEAR RISE OVER 17 WEEKS BEFORE LEAD WAS REVISED. THRESHOLDS: THRESHOLDS INCREASE OVER 16 WEEKS, STARTING 27-MAY, TOPPING OUT AT 1V. WITHIN NORMAL RANGE.
IT WAS REPORTED THAT THE LEAD EXHIBITED SLOWLY INCREASING THRESHOLDS AND IMPEDANCES, REACHING HIGH LEVELS OF IMPEDANCE. IT WAS FURTHER REPORTED THAT THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE LEAD EXHIBITED SLOWLY INCREASING THRESHOLDS AND IMPEDANCES, REACHING HIGH LEVELS OF IMPEDANCE. IT WAS FURTHER REPORTED THAT THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE SENSE | ELECTRODE, PACEMAKER, PERMANENT | DTB | MEDTRONIC PUERTO RICO, INC. | 4074 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |