FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR AT

MDR report key: 2810598 · Received October 31, 2012

Report

Report Number
6000094-2012-02415
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC S.A.
Product Code
NIK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. EVALUATION SUMMARY (B)(4) - WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. BATTERY DEPLETION WAS INDICATED. TIME OF ELECTIVE REPLACEMENT INDICATOR IN SAVE TO DISK WAS ON (B)(6) 2012. DEVICE ELECTIVE REPLACEMENT INDICATOR WAS LESS THAN OR EQUAL TO 2.62 VOLTS. THE WEEKLY BATTERY VOLTAGE TREND DATA SHOWS MINIMUM BATTERY VOLTAGE EQUAL TO 2.64 TO 2.62 VOLTS BETWEEN (B)(6) 2012. THERE WAS ONE PATIENT ALERT FOR LOW BATTERY VOLTAGE ON (B)(6) 2012. (B)(4) - THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 80% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) FASTER THAN EXPECTED. THE DEVICE LONGEVITY MEASURED FIVE DAYS BEFORE ERI TRIGGERED WAS 2.93 VOLTS. HOWEVER, THE BATTERY VOLTAGE MEASURED SIX MONTHS PRIOR TO THAT INDICATED 2.74 VOLTS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR AT IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC S.A. C174AWK

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R