FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 2810597
·
Received October 31, 2012
Report
- Report Number
- 2649622-2012-16017
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- September 14, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THE OUTER INSULATION WAS BREACHED CUT, EXHIBITED A COSMETIC CUT, AND THE LEAD APPEARED TO HAVE BEEN DAMAGED AT IMPLANT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE LEAD IMPLANT PROCEDURE, THE PHYSICIAN ATTEMPTED TO EXTEND THE HELIX OUTSIDE THE BODY PRIOR TO THE LEAD IMPLANT; HOWEVER, THE HELIX WOULD NOT EXTEND AFTER MORE THAN 20 TURNS. THE LEAD WAS NOT USED AND A DIFFERENT LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Other |