FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2810597 · Received October 31, 2012

Report

Report Number
2649622-2012-16017
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
September 14, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THE OUTER INSULATION WAS BREACHED CUT, EXHIBITED A COSMETIC CUT, AND THE LEAD APPEARED TO HAVE BEEN DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE LEAD IMPLANT PROCEDURE, THE PHYSICIAN ATTEMPTED TO EXTEND THE HELIX OUTSIDE THE BODY PRIOR TO THE LEAD IMPLANT; HOWEVER, THE HELIX WOULD NOT EXTEND AFTER MORE THAN 20 TURNS. THE LEAD WAS NOT USED AND A DIFFERENT LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076

Patients

Seq Age Sex Outcome Treatment
1 82 YR Other