VIRTUOSO DR
Report
- Report Number
- 6000144-2012-05893
- Event Type
- Injury
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S062
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.
THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED AND REPLACED DUE TO ERI (ELECTIVE REPLACEMENT INDICATOR) WITHIN FIVE YEARS OF IMPLANT. THE DEVICE WAS ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VIRTUOSO DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D154AWG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD |