FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 2810513 · Received October 31, 2012

Report

Report Number
6000144-2012-06029
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYSIS REVEALED NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD POSSIBLE EARLY BATTERY DEPLETION. IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD VERY HIGH THRESHOLDS. BOTH THE DEVICE AND THE LV LEAD WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSULTA CRT-D IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. D224TRK

Patients

Seq Age Sex Outcome Treatment
1 86 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD