FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2810485 · Received October 31, 2012

Report

Report Number
2649622-2012-15743
Event Type
Injury
Date Received
October 31, 2012
Report Date
August 11, 2014
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S17
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS ALSO REPORTED THAT THE LEAD WAS EXPLANTED AND REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS AN ALERT FOR T-WAVE OVERSENSING, NOISE AND HIGH IMPEDANCE ON THE RIGHT VENTRICULAR LEAD. THE LEAD ALSO HAD A POSSIBLE FRACTURE. THE LEAD WAS REPROGRAMMED, IS STILL IN USE AND WILL BE MONITORED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC PUERTO RICO, INC. 6944

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| 4193 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB