FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 2810481 · Received October 31, 2012

Report

Report Number
6000094-2012-02366
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC S.A.
Product Code
NIK
Removal / Correction Number
Z-0111-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE INDICATED RECOMMENDED REPLACEMENT TIME (RRT) AND THE DEVICE HAD UNEXPECTED LONGEVITY. THE DEVICE WILL BE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSULTA CRT-D IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC S.A. D234TRK

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R