FDA Adverse Event Malfunction Summary report: N

SPRINT FIDELIS

MDR report key: 2810428 · Received October 31, 2012

Report

Report Number
2649622-2012-15730
Event Type
Malfunction
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S029
Removal / Correction Number
Z-0070-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) - THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD SYNCOPAL EPISODES. THE REMOTE MONITOR SHOWED EVIDENCE OF PAUSES IN VENTRICULAR PACING. IT WAS NOTED THAT THE VENTRICULAR SENSE RESPONSE ALGORITHM DID NOT CONTINUOUSLY PACE THE VENTRICLE DURING THE VENTRICULAR OVERSENSING EPISODES. BOTH THE DEVICE AND THE RIGHT VENTRICULAR LEAD ARE STILL IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT FIDELIS IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6949

Patients

Seq Age Sex Outcome Treatment
1 80 YR Life Threatening| O 1388T X 2 COMPETITOR IMPLANTABLE PACING LEADS