FDA Adverse Event
Malfunction
Summary report: N
OT ULTRALINK METER
MDR report key: 2810362
·
Received October 31, 2012
Report
- Report Number
- 2939301-2012-12523
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Report Date
- October 17, 2012
- Manufacturer
- LIFESCAN INC.
- Product Code
- NBW
- PMA / PMN Number
- K073231
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4) - DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED ON (B)(4) 2012 AND EVALUATED BY PRODUCT ANALYSIS (PA) ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE METER PASSED TESTING WITH NO FAULTS FOUND. THE METER WAS FOUND TO FUNCTION PROPERLY. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
Description of Event or Problem · 1
THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE ERROR 4 WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT ULTRALINK METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN INC. | 3339214 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |