FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2810335 · Received October 31, 2012

Report

Report Number
2032227-2012-07597
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 16, 2012
Report Date
October 16, 2012
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

MOTOR ERROR ALARMED DURING REWIND DUE TO CORROSION ON MOTOR HOME SWITCH. CORROSION FOUND ON TRANSDUCER AND INTERFACE BOARD. UNABLE TO VERIFY SEVERAL MOTOR ERROR ALARMS DUE TO MOTOR DAMAGE. KEYPAD BUTTON AND DISPLAY FUNCTION PROPERLY. NO FROZEN DISPLAY OR DAMAGE FOUND ON KEYPAD TRACE.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME

Description of Event or Problem · 1

THE CUSTOMER REPORTED A FROZEN DISPLAY AND UNRESPONSIVE BUTTONS ON THE INSULIN PUMP. THE CUSTOMER STATED THAT THE TIME ON THE SCREEN WAS NOT ADVANCING EITHER. TROUBLESHOOTING WAS PERFORMED, BUT THE ISSUES WERE NOT RESOLVED. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP OYC OYC MEDTRONIC MINIMED MMT-722LNAB

Patients

Seq Age Sex Outcome Treatment
1 34 YR