FDA Adverse Event Injury Summary report: N

SROM STM ST,36+6L NK,11X16X150

MDR report key: 2810242 · Received October 31, 2012

Report

Report Number
1818910-2012-26695
Event Type
Injury
Date Received
October 31, 2012
Date of Event
January 3, 2007
Report Date
October 17, 2012
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
MRA
PMA / PMN Number
P040023
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION SINCE ITS RELEASE FOR DISTRIBUTION. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE PATIENTS REVISION SURGERY OPERATIVE NOTES WERE PROVIDED AND REVIEWED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE FOR THE NEWLY ADDED PRODUCTS AS THE REQUIRED PRODUCT AND LOT CODES WERE NOT PROVIDED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. BASED ON THE INABILITY TO IDENTIFY A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). UDI: (B)(4).

Description of Event or Problem · 1

CLINICAL REPORT STATES THE PATIENT WAS REVISED FOR ASEPTIC LOOSENING OF THE STEM. (RIGHT HIP).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SROM STM ST,36+6L NK,11X16X150 FEMORAL STEM MRA DEPUY ORTHOPAEDICS, INC. 1818910 2015112

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention