FREESTYLE LITE
Report
- Report Number
- 2954323-2012-06684
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- October 7, 2012
- Report Date
- October 7, 2012
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED. (B)(4).
AS PRODUCT WAS NOT RETURNED AND NO TEST STRIP LOT WAS REPORTED WITH THIS COMPLAINT, A DEVICE HISTORY REVIEW (DHR) OF THE METER WAS REQUESTED. THE DHR FOR METER SERIAL NUMBER (B)(4) INDICATED THE DEVICE WAS PERFORMING WITHIN ITS PERFORMANCE CLAIMS AND MET THE MANUFACTURER'S QUALITY SPECIFICATIONS PRIOR TO ITS RELEASE. THIS IS A FINAL REPORT.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 87 MG/DL AND 417 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 MO |