FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 2810222 · Received October 31, 2012

Report

Report Number
2954323-2012-06684
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 7, 2012
Report Date
October 7, 2012
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

AS PRODUCT WAS NOT RETURNED AND NO TEST STRIP LOT WAS REPORTED WITH THIS COMPLAINT, A DEVICE HISTORY REVIEW (DHR) OF THE METER WAS REQUESTED. THE DHR FOR METER SERIAL NUMBER (B)(4) INDICATED THE DEVICE WAS PERFORMING WITHIN ITS PERFORMANCE CLAIMS AND MET THE MANUFACTURER'S QUALITY SPECIFICATIONS PRIOR TO ITS RELEASE. THIS IS A FINAL REPORT.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 87 MG/DL AND 417 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE LITE BLOOD GLUCOSE MONITORING SYSTEM NBW

Patients

Seq Age Sex Outcome Treatment
1 33 MO