FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2810201 · Received October 31, 2012

Report

Report Number
2531779-2012-12959
Event Type
Malfunction
Date Received
October 31, 2012
Report Date
October 18, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4): THE KEYPAD WAS FOUND TO BE FULLY INTACT; NO PEELING OR DAMAGE WAS OBSERVED. DURING TESTING, THE UP ARROW KEYPAD BUTTON WAS INTERMITTENTLY UNRESPONSIVE; ALL OTHER BUTTONS RESPONDED APPROPRIATELY. DURING INVESTIGATION, EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

ON (B)(6) 2012, THE PATIENT CONTACTED ANIMAS AND REPORTED THAT THE UP ARROW KEYPAD BUTTON WAS DELAYED AND INTERMITTENTLY UNRESPONSIVE. THE PATIENT DENIED ANY DAMAGE TO THE RUBBER KEYPAD OR EVIDENCE OF MOISTURE IN THE PUMP. THE PATIENT REPORTEDLY WORE THE PUMP IN A POCKET AND DID NOT CLEAN THE PUMP. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 45 YR