FDA Adverse Event Other Summary report: N

VAPOTHERM DISPOSABLE PT CIRCUIT

MDR report key: 2810200 · Received October 3, 2012

Report

Report Number
1125759-2012-00003
Event Type
Other
Date Received
October 3, 2012
Date of Event
April 4, 2012
Report Date
October 3, 2012
Manufacturer
VAPOTHERM, INC
Product Code
BTT
PMA / PMN Number
K072845
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

MANUFACTURER'S RESPONSE: WE RECEIVED A REPORT FROM THE CUSTOMER REGARDING WATER COMING THROUGH THE CANNULA. WE INVESTIGATED THE DISPOSABLE PT CIRCUIT AND FOUND THAT THE LEAK RATE WAS 3 MILLILITERS/HOUR. WE WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE APPROPRIATE ACTION AS NECESSARY. EVENT DESCRIPTION: PT WAS ON THE VAPOTHERM HIGH FLOW NASAL CANNULA RECEIVING CARE FROM PT WHEN IT WAS NOTICED THAT WATER WAS COMING THROUGH THE CANNULA AND SPRAYING INTO THE PT'S NOSE. THE VAPOTHERM WAS REMOVED FROM THE PT AND INSPECTED BY RT. IT WAS FOUND TO BE PERIODICALLY SPRAYING WATER OUT OF THE CIRCUIT WHERE THE CANNULA CONNECTS. PT WAS PLACED ON A DIFFERENT NASAL CANNULA AND THE CIRCUIT WAS REMOVED FORM THE CAPOTHERM AND THE RT SUPERVISOR WAS NOTIFIED OF THE INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VAPOTHERM DISPOSABLE PT CIRCUIT PF-DPC-LOW BTT VAPOTHERM, INC PF-DPC-LOW LF1202007

Patients

Seq Age Sex Outcome Treatment
1 5 MO Other