LIGHTSHEER XC
Report
- Report Number
- 1720381-2012-00058
- Event Type
- Injury
- Date Received
- October 31, 2012
- Date of Event
- September 8, 2012
- Report Date
- October 31, 2012
- Manufacturer
- RH USA, INC.
- Product Code
- GEX
- PMA / PMN Number
- K003614
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
LUMENIS INVESTIGATED THE EVENT REPORT CONTACTING THE USER FACILITY TO REQUEST PATIENT TREATMENT SETTINGS AND PATIENT PHOTOGRAPHS. THE PATIENT TREATMENT SETTINGS WERE PROVIDED BY THE FACILITY; HOWEVER NO PATIENT PHOTOGRAPHS WERE AVAILABLE. THE USER FACILITY DECLINED SERVICE OF THE SUBJECT DEVICE BY LUMENIS TECHNICAL REPRESENTATIVE. LUMENIS HAS NOT HELD A SERVICE CONTRACT WITH THE USER FACILITY SINCE 07/30/2007. THE USER FACILITY PURCHASED A NEW HANDPIECE ON (B)(4) 2012. A LUMENIS TECHNICAL ENGINEER COMPLETED PERFORMANCE TESTS OF ALL SPECIFICATIONS CONCLUDING THAT THE DEVICE OPERATED TO MANUFACTURER'S SPECIFICATIONS AT THAT TIME. LUMENIS CANNOT RULE OUT THAT SERVICE BY UNAUTHORIZED THIRD PARTY SERVICE PROVIDERS MAY HAVE CONTRIBUTED TO THE REPORTED ADVERSE EVENT. TO THE BEST OF LUMENIS' KNOWLEDGE, THE SUBJECT DEVICE REMAINS IN USE AT THE USER FACILITY. A LUMENIS HEALTHCARE PROFESSIONAL EVALUATED THE REPORTED TREATMENT SETTINGS CONCLUDING THAT THE REPORTED SETTINGS WERE WITHIN NORMAL LIMITS PER DEVICE TRAINING AND DEVICE IFU, AND THAT THE INFORMATION PROVIDED BY THE USER FACILITY WAS INSUFFICIENT TO MAKE A DETERMINATION OF CAUSE. USER FACILITY DECLINED SERVICE.
IT WAS REPORTED BY A USER FACILITY THAT A PATIENT OF SKIN TYPE VI SUSTAINED A SECOND DEGREE BURN TO THE GENITAL AREA FOLLOWING A TEST PATCH WITH A LUMENIS LIGHTSHEER XC LASER. NO INFORMATION REGARDING MEDICAL INTERVENTION TO PRECLUDE SERIOUS INJURY WAS RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGHTSHEER XC | SOLID STATE AESTHETIC LASER DELIVERY DEVICE | GEX | RH USA, INC. | LIGHTSHEER XC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR | Other |