FDA Adverse Event
Malfunction
Summary report: N
ENDOLAP
MDR report key: 280907
·
Received June 8, 2000
Report
- Report Number
- MW1019080
- Event Type
- Malfunction
- Date Received
- June 8, 2000
- Date of Event
- June 8, 2000
- Report Date
- June 8, 2000
- Manufacturer
- ENDOLAP, INC
- Product Code
- EQH
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- OR, US
- Reporter Occupation
- PHARMACIST
Narratives
Description of Event or Problem · 1
DELIVERY END OF THE CORD IS A LIGHT CARRIER AND IT STICKS OUT OVER THE COUPLER END BY ABOUT 3-4MM. THE LIGHT CORD WAS LAID ON A TOWEL IN A STERILE FIELD AND WHEN IT WAS TURNED ON, WITHIN SECONDS, THE TOWEL CAUGHT FIRE. RPTR STATES THAT A COMPETITOR MAKES THE SAME TYPE OF CORD WHICH HAS A COVER GUARD FOR THE END. RPTR BELIEVES THAT IF DEVICE HAD THIS GUARD THE FIRE WOULD NOT HAVE STARTED. MFR WAS NOTIFIED ABOUT EVENT AND IS INVESTIGATING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOLAP | FIBEROPTIC LIGHT CORD 0.4MM X 10FT | EQH | ENDOLAP, INC | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |