FDA Adverse Event Malfunction Summary report: N

GLUCOMETER ELITE 4 BASIC

MDR report key: 280874 · Received May 25, 2000

Report

Report Number
1810909-2000-00020
Event Type
Malfunction
Date Received
May 25, 2000
Date of Event
January 19, 2000
Report Date
May 24, 2000
Manufacturer
KDK CORPORATION
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

THE COMPLAINT INDICATED THAT THE GLUCOMETER ELITE 4 BASIC IS READING MUCH LOWER WHEN COMPARED TO ANOTHER METHOD. COMPLAINANT RECEIVING A RESULT IN THE 30 MG/DL RANGE WHILE A COMPETITIVE METHOD RAN AT APPROX THE SAME TIME GAVE A RESULT OVER 300 MG/DL. (EXACT VALUE UNKNOWN). WHILE ON THE PHONE, IT WAS LEARNED THAT THE CUSTOMER WAS USING EXCEL OFF BRAND REAGENT STRIPS. IT WAS EXPLAINED TO THE CUSTOMER THAT THIS OFF BRAND REAGENT WAS NOT SUPPLIED BY BAYER AND IT IS UNKNOWN HOW THESE OFF BRAND REAGENT STRIPS WORK IN A SYSTEM THAT WAS DESIGNED TO BE USED WITH ELITE SENSORS ONLY. REPLACEMENT REAGENT WAS PROVIDED TO THE CUSTOMER AND CONTROL TESTING, AS WELL AS ELECTRONIC CHECKS WERE SATISFACTORY. THE CUSTOMER WAS ADVISED TO CONTACT THE MFR OF THE EXCEL OFF BRAND REAGENT FOR FURTHER INVESTIGATION. THE COMPLAINT IS CONSIDERED CLOSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GLUCOMETER ELITE 4 BASIC BLOOD GLUCOSE METER NBW KDK CORPORATION 3901N *

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN