FDA Adverse Event Malfunction Summary report: N

GCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN

MDR report key: 2805490 · Received September 22, 2008

Report

Report Number
3003900188-2008-00042
Event Type
Malfunction
Date Received
September 22, 2008
Manufacturer
LINDE GAS NORTH AMERICA LLC
Product Code
ECX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRESENT CASE IS ONE OUT OF 9 INCIDENT REPORTS IN CONNECTION WITH 7B LIV IGNITIONS. EVALUATION SUMMARY: ROOT CAUSE ANALYSIS: THE CAUSE OF IGNITION WITH HIGHEST PROBABILITY IS IGNITION OF HYDROCARBONS, LUBRICANT AND/OR DEBRIS BY ADIABATIC COMPRESSION. POTENTIAL SOURCES OF HYDROCARBONS AND LUBRICANTS; LUBRICANT (FOMBLIN) MIGRATION (LUBRICANT FROM GCE MANUFACTURING OF VALVE). LEACHING OF CONTAMINANTS (PHTHALATES) FROM O-RINGS IN VALVES (LUBRICANT IN SOLVENT) AND HYDROCARBON OIL. LEACHING OF CONTAMINANTS (PHTHALATES) FROM FILL CONNECTOR AT FILLING PLANT AT LINDE. POTENTIAL SOURCES OF DEBRIS. DEBRIS FROM MANUFACTURING PROCESS OF VALVE AT GCE. DEBRIS FROM FILLING PLANT AT LINDE. DEBRIS FROM WEAR, VALVES IN USE. THE DESIGN (FILTER AND LOCATION OF COMPONENTS IN THE HIGH PRESSURE SIDE OF VALVE) OF THE GCE LIV VALVE 5 MICRON FILTER IS LIKELY TO CAUSE MORE SENSITIVITY TO ADIABATIC COMPRESSION HEATING THAN THE GCE OXYGEN DESIGN.

Description of Event or Problem · 1

ON (B)(6) 2008, LINDE WAS INITIALLY INFORMED THAT AN INCIDENT INVOLVING LIV MAXIFLOW 2 LITERS HAD OCCURRED IN THE (B)(6) 2008. NO PEOPLE WERE INJURED BY THE ACCIDENT. AN AMBULANCE IN (B)(6), WAS SENT TO AN INCIDENT WHERE A PATIENT WAS LYING ON A TERRACE. ONE OF THE PARAMEDICS WANTED TO GIVE MEDICINAL OXYGEN TREATMENT TO THE PATIENT AND CONNECTED A HOSE TO THE HOSE BARB OF THE LIV CYLINDER IN HORIZONTAL POSITION. SHORTLY AFTER OPENING THE MAIN VALVE OF THE CONOXIA 2 LITER LIV CYLINDER AND SETTING THE FLOW-REGULATOR TO 12 L/MIN, A BIG BANG WAS HEARD AND A FLAME CAME OUT OF THE CONOXIA CYLINDER. THE PARAMEDIC CLOSED THE MAIN SHUT OFF VALVE IMMEDIATELY AND DISCONNECTED THE PATIENT. A POLICEMAN AT THE SCENE EXTINGUISHED THE FIRE (ACCORDING TO SPECTATORS). THERE WAS SOOT INSIDE THE HOSE AND MASK AND IT IS UNKNOWN IF THE PATIENT INHALED ANY OF THE SOOT. THERE WERE NO INJURIES AS A RESULT OF THE IGNITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GCE COMBILITE I VIPRS, 5 MICRON FILTER, LIV FOR OXYGEN NONE ECX LINDE GAS NORTH AMERICA LLC 0710204 2657674

Patients

Seq Age Sex Outcome Treatment
1